• prduct1

Products

ET1625High Precision Multifunction Process Calibrator

Short Description:

ET1625 high-precision multi-function process calibrator is a high-precision, hand-held process signal measurement/output instrument, which can simultaneously measure and output voltage, current, resistance, thermocouple, thermal resistance, frequency, pulse, switch value, etc. Industrial process signal; built-in HART function, can completely replace the HART Communicator.

The product can replace the measurement and calibration instruments such as current signal source, voltage signal source, resistance box, electronic potentiometer, frequency meter, HART hand communicator, etc. It is mainly used in industrial field signal calibration and fault diagnosis; it is also suitable for signal measurement and calibration in chemical industry, military industry and various research institutes and laboratories. This product is a powerful, standard-grade industrial process detection instrument for field and laboratory use.


Product Detail

Product Tags

Features

●Measurement/output: voltage, current, resistance, frequency, pulse, switch value, among which current output supports active and passive;
●Simulate thermal resistance and thermocouple in the form of temperature;
●Can simulate two-wire transmitter;
●Resistance measurement can be selected in two, three and four-wire ways;
●Accuracy is 0.01%;
●Two mutually isolated channels, support synchronous measurement and output;
●Can provide manual step, automatic step, automatic step and manual step function;
●Using 3.5-inch TFT LCD screen with a resolution of 480*320;
●The measurement and output data can be displayed at the same time, and the measurement or output data can also be displayed separately;
●5000mAh lithium battery power supply;
●With automatic power off function, time can be set, suitable for on-site use;
●Provide DC24V loop power supply for on-site debugging;
●The thermocouple measurement and output provide built-in, external and manual cold junction temperature compensation methods, of which the external reference terminal adopts A-grade Pt100 platinum resistance;
●Thermocouple Type:R,S,K,E,J,T,N,B,L,U,XK,WRE325,WRE526;
●Type of thermal resistance: PT100-385,PT100-392,PT100-JIS,PT200-385,PT500-385,PT1000-385,
Cu10,Cu50,Cu100,Nil20,BA1,BA2,PT10;
HART function (only ET1625H and ET1615H support this function): It can completely replace the HART hand-held communicator, set or calibrate the range of the smart transmitter, and force the output current of the smart transmitter to be at a fixed value (such as 20mA, 12mA) , 4mA), set the linear or square root function, can perform HART reset on the pressure sensor of the transmitter, etc.;

Technical indicators

Model

ET1625

ET1625H

ET1615

ET1615H

ET1610

notes

Voltage output

range

100mV

Resolution

0.1µV

Maximum load current<=2.5mA

precision

0.008%RD+0.003%FS

1V

Resolution

1µV

precision

0.008%RD+0.001%FS

10V

Resolution

10µV

precision

0.008%RD+0.001%FS

Current output (active and passive)

30mA

Resolution

0.1µA

Active output maximum load voltage20V

precision

0.008%RD+0.003%FS

Frequency output

range

100Hz

Resolution

0.001Hz

Maximum load current≤2.5mA

precision

0.01%FS

1kHz

Resolution

0.01Hz

precision

0.01%FS

10kHz

Resolution

10Hz

precision

0.01%FS

Resistive output

range

50Ω

Resolution

0.1mΩ

Excitation current0.4-4mA

precision

0.008%RD+30 mΩ

500Ω

Resolution

1mΩ

Excitation current0.1-2mA

precision

0.008%RD+30 mΩ

5000Ω

Resolution

10mΩ

Excitation current0.04-0.4mA

precision

0.008%RD+100 mΩ

Pulse output

range

10Hz(1~100000)

1cyc

±2dig

Maximum load current≤2.5mA

1kHz(1~100000)

100kHz(1~100000)

Switch output

range

100Hz(1Hz~110Hz)

Resolution

0.01Hz

precision

±2dig

1kHz(0.1kHz~1.1kHz)

Resolution

1Hz

precision

±2dig

10kHz(1kHz~11kHz)

Resolution

0.1kHz

precision

±2dig

100kHz(10kHz~110kHz)

Resolution

1kHz

precision

±2dig

24output

24V

±10%FS

loop output

model

ET1625

ET1625H

ET1615

ET1615H

ET1610

note

RTD output

See the thermal resistance table for details

See the thermal resistance table for details

Thermocouple output

See thermocouple table for details

See thermocouple table for details

Voltage measurement

range

200mV

Resolution

0.1µV

precision

0.008%RD+0.003%FS

2V

Resolution

1µV

precision

0.008%RD+0.002%FS

20V

Resolution

10µV

precision

0.008%RD+0.002%FS

200V

Resolution

0.1mV

precision

0.008%RD+0.002%FS

Current measurement range

20mA

Resolution

0.1µA

precision

0.008%RD+0.003%FS

200mA

Resolution

1µA

precision

0.008%RD+0.003%FS

Frequency measurement

range

100Hz

Resolution

0.001Hz

Maximum load current≤2.5mA

precision

0.01%FS

1kHz

Resolution

0.01Hz

precision

0.01%FS

10kHz

Resolution

10Hz

precision

0.01%FS

Resistance measurement (four-wire)

range

50Ω

Resolution

0.1mΩ

Excitation current1mA

precision

0.008%RD+30 mΩ

500Ω

Resolution

1mΩ

precision

0.008%RD+30 mΩ

5000Ω

Resolution

10mΩ

Excitation current0.1mA

precision

0.008%RD+100mΩ

Resistance measurement (two or three wires)

range

50Ω

Resolution

0.1mΩ

0.1mΩ

precision

0.008%RD+30 mΩ(3 line);0.008%RD+80 mΩ(2 line)

0.008%RD+30 mΩ(3 line);0.008%RD+80 mΩ(2 line)

500Ω

Resolution

1mΩ

1mΩ

precision

0.008%RD+50 mΩ(3线);0.008%RD+80 mΩ(2 line)

0.008%RD+50 mΩ(3 line);0.008%RD+80 mΩ(2 line)

5000Ω

Resolution

10mΩ

10mΩ

precision

0.008%RD+200mΩ

0.008%RD+200mΩ

Model

ET1625

ET1625H

ET1615

ET1615H

ET1610

note

Switch measurement

CLOSE/OPEN

Excitation current1mA

Thermal resistance measurement

See the thermal resistance table for details

See the thermal resistance table for details

Thermocouple Measurement

See thermocouple table for details

See thermocouple table for details

HART

have

have

Thermal resistance table

Graduation number

temperature range

Resolution

Accuracy

note

PT10

-200-850℃

0.01℃

0.2℃

Four-wire measurement

PT100-385

-200-850℃

0.01℃

0.1℃

PT100-392

-200-850℃

0.01℃

0.1℃

PT100-JIS

-200-850℃

0.01℃

0.1℃

PT200-385

-200-630℃

0.01℃

0.1℃

PT500-385

-200-630℃

0.01℃

0.2℃

PT1000-385

-200-650℃

0.01℃

0.1℃

Cu10

-100-260℃

0.01℃

0.5℃

Cu50

-50-150℃

0.01℃

0.2℃

Cu100

-50-150℃

0.01℃

0.2℃

BA1

-200-650℃

0.01℃

0.4℃

BA2

-200-650℃

0.01℃

0.25℃

Ni20

-80-260℃

0.01℃

0.3℃

Thermocouple table

Graduation number

temperature range

Resolution

Accuracy

note

K

-200-0℃

0-1372℃

0.1℃

0.4℃

0.3℃

not include cold junction compensation accuracy

R

-50-0℃

0-1768℃

0.1℃

0.9℃

0.7℃

S

-50-0℃

0-1768℃

0.1℃

0.9℃

0.6℃

E

-50-0℃

0-1000℃

0.1℃

0.5℃

0.4℃

J

-200-0℃

0-1200℃

0.1℃

0.3℃

0.2℃

T

-100-0℃

0-400℃

0.1℃

0.6℃

0.3℃

L

-200-900℃

0.1℃

0.2℃

N

-200-0℃

0-1300℃

0.1℃

0.7℃

0.5℃

B

600-1820℃

0.1℃

0.7℃

U

-200-0℃

0-400℃

0.1℃

0.6℃

0.4℃

XK

-200-800℃

0.1℃

0.5℃

WRE325

0-1500℃

0.1℃

0.6℃

WRE526

0-1500℃

0.1℃

0.6℃

Note: The built-in cold junction compensation accuracy is 0.5℃

 

    -100~0℃

±0.5℃

      0~400℃

±0.2℃

Note: The temperature of the reference end is 0℃; the error of the internal temperature probe: ±0.2℃, following the ITS-90 international temperature scale.

General technical specifications

Use environment: temperature 20℃±2℃, relative humidity≦80%RH
Dimensions of the calibration table: 1880mm (length) × 758mm (width) × 1460mm (height)
Power supply: AC220V±10%, 50Hz±2Hz
Power consumption: <500W


  • Previous:
  • Next:

  • Write your message here and send it to us